Clinical-Translational Studies of Diagnostic Devices

Biostatistics Services

Biostatistics Services for IVD and AI Clinical-Translational Studies

Anatomise Biostats provides specialised biostatistics support for clinical-translational studies of In Vitro Diagnostics (IVDs) and AI-driven medical devices. Targeted statistical methodologies bridge the gap between early biomarker discovery and late-stage clinical trials for liquid biopsies, ultra-rapid point-of-care (POC) blood tests, and AI algorithms (diagnostic, prognostic, or disease monitoring), supporting the transition from translational research to regulatory submission.

Schedule a Discovery Call
Ultra-rapid point-of-care blood testing device
The Role

The Role of Biostatistics in Clinical-Translational Studies

Translating novel biomarkers or AI models into clinical practice requires precise data analysis to establish analytical and clinical validity. Our biostatisticians understand the specific statistical methods needed to interpret complex translational data, such as handling low variant allele frequencies (VAF) in liquid biopsy, evaluating turnaround times (TAT) in POC testing, or assessing model overfitting and calibration in machine learning algorithms. Partnering with an independent statistical team provides objective validation, lending credibility to the clinical-translational study data.

Study Design

Clinical-Translational Study Design for IVDs and AI Algorithms

Diagnostic devices and their underlying AI/ML models require tailored study designs to evaluate performance against a reference standard. Study protocols are developed in collaboration with your team, aligned with CLSI guidelines, the FDA AI/ML SaMD action plan, and MHRA expectations. The approach includes:

Retrospective Biobank and Prospective Cohorts

Utilising retrospective biobank specimens for early biomarker discovery and algorithm training, followed by prospective translational cohorts to validate clinical performance and lock the algorithm.

Reference Standard Selection

Establishing the appropriate composite reference standard (CRS) for comparison. In liquid biopsy validation, where tissue biopsies may suffer from tumour heterogeneity, latent class analysis is implemented to account for an imperfect gold standard, with extended models utilised when conditional independence assumptions are violated.

Bias Mitigation

Designing studies to avoid spectrum bias and verification bias. For AI models, selection bias in training and tuning datasets is minimised to support algorithmic generalisability across diverse demographic subgroups.

Data Partitioning for AI

Establishing predefined training, validation, and independent test (holdout) datasets. Strict separation of these datasets is maintained to prevent data leakage, a critical requirement for validating locked algorithms and supporting Predetermined Change Control Plans (PCCPs) that govern bounded, pre-specified post-market modifications.

Resource Efficiency

For diagnostic and AI validation studies, precision-based sample size calculations are often more relevant than power-based calculations. Sample sizes are calculated to achieve narrow 95% confidence intervals for sensitivity and specificity, preventing underpowered studies or unnecessary patient enrolment.

Validation

Biomarker Discovery and AI Algorithm Validation

Validating diagnostic accuracy and algorithm performance requires specific statistical frameworks tailored to the technology. Advanced methods evaluate the performance of liquid biopsies, POC blood tests, and AI-driven software (SaMD):

High-Dimensional Data and Feature Selection

For multi-omics or complex biomarker discovery, False Discovery Rate (FDR) is controlled using the Benjamini-Hochberg procedure. Regularised regression techniques (LASSO, Elastic Net) are applied for feature selection to prevent model overfitting in high-dimensional translational datasets.

AI Model Calibration and Discrimination

For diagnostic and prognostic algorithms, calibration is assessed using calibration plots, Brier scores, and the Hosmer-Lemeshow test. Discrimination is evaluated using the Area Under the Curve (AUC) of the Receiver Operating Characteristic (ROC) or Precision-Recall (PR) curves, particularly for imbalanced datasets common in rare mutation detection.

Algorithmic Locking and Drift

Nested cross-validation is utilised during the training phase to tune hyperparameters without leaking test data. For disease monitoring algorithms, time-series analyses and longitudinal mixed-effects models are applied to track disease progression and evaluate algorithmic drift over time.

IVD Analytical Validation

Clinical validation is supported by the statistical analysis of analytical performance, including Limit of Blank (LoB), Limit of Detection (LoD), and Limit of Quantitation (LoQ) per CLSI EP17 guidelines. For continuous biomarkers, Passing-Bablok and Deming regression are utilised to assess agreement with the reference method.

Clinical Utility

Demonstrating Clinical Utility in Translational Research

Moving beyond analytical and clinical validity requires demonstrating clinical utility. Decision Curve Analysis (DCA) is utilised to quantify the net benefit of the diagnostic or prognostic algorithm across different threshold probabilities. Additionally, Net Reclassification Improvement (NRI) and Integrated Discrimination Improvement (IDI) are calculated to demonstrate how the novel algorithm improves risk stratification compared to standard clinical models. This provides complementary reclassification metrics to support the real-world impact of the device for payers and providers.

Real-World Evidence

Real-World Evidence and Post-Market Surveillance

Demonstrating long-term clinical utility requires data beyond the controlled clinical trial setting. Analysis of real-world evidence (RWE) shows how the device or algorithm performs across diverse patient populations and healthcare settings. By analysing observational data and post-market surveillance metrics, the impact on patient management is defined. This includes calculating Number Needed to Screen (NNS), evaluating shifts in PPV/NPV based on real-world disease prevalence, and monitoring for algorithmic performance drift over time.

Regulatory

Regulatory Compliance and Submissions

Meeting regulatory requirements for IVDs and AI SaMD requires precise statistical documentation. The statistical sections for regulatory submissions are prepared, including Clinical Evaluation Reports (CER) for EU MDR, Performance Evaluation Reports (PER) for EU IVDR, as well as 510(k) and De Novo applications for the FDA.

Statistical Analysis Plans (SAPs)

Detailed SAPs are drafted to lock the analysis methodology before unblinding the data.

Agency Interactions

Direct statistical support is provided during FDA Pre-Submission (Pre-Sub) meetings or Notified Body reviews, defending methodology choices and addressing reviewer queries.

Data Quality and Interpretation

Clear separation between the sponsor and independent analysts is maintained during data analysis to prevent confirmation bias, while collaborating closely on study design and SAP finalisation prior to unblinding. Data collection is structured using CDASH and SDTM standards, and ALCOA+ principles are implemented for data management. Handling indeterminate or invalid results is a critical component of diagnostic study analysis. Predefined statistical rules are applied for handling these missing or uninterpretable data points without introducing bias, providing clear data visualisation and clinical study reports (CSRs) that translate complex statistical findings into actionable clinical insights.

Why Work With Anatomise Biostats?

Direct collaboration with your clinical and engineering teams allows the study design to fit the specific risk profile and intended use of your IVD or AI algorithm. From early translational biomarker discovery to post-market clinical follow-up (PMCF), the objective is to produce statistically sound evidence that supports regulatory approval and market adoption.